| 1 | Mobile unit | 1 | ||
| Type | : | Mobile X-ray unit with telescopic X-ray tube arm | ||
| Dimensions (l x w x h) in parking position | ± 1382 mm x 670 mm x 1330 mm (54.4″ x 26.4″ x 52.4″) | |||
| Wheel base length | : | ± 600 mm (23.6″) | ||
| Motorization (speed range) | : | Max 5 km/h (3.1 mph) tube in parking position | ||
| Focal point distance from floor | : | ± 53 to 202 cm (20.9″ to 79.5″) | ||
| Focal point distance to column | : | min: 702 mm; max: 1242 mm (49″) | ||
| Tube column rotation | : | ± 317° | ||
| Batteries | : | Separate batteries for drive and generator control | ||
| Power for charging | : | single phase 230/115 VAC ± 10%; 50/60Hz | ||
| Fine positioning from tube head | : | Yes | ||
| Anti-collision sensor and brake | : | Yes | ||
| 2 | GENERATOR | 1 | ||
| Power | : | 40 kW | ||
| Tube voltage in steps of 1 kV | : | 40 – 150 kV in steps of 1 kV | ||
| mA range | : | 10 – 500 mA | ||
| mAs range | : | 0.1 – 500 mAs | ||
| 3 | X-RAY TUBE | 1 | ||
| Type | : | Rotating Anode | ||
| Focal spot (small/large) | : | 0.7 / 1.3 | ||
| Anode heat storage capability | : | 220 kJ (300 kHU) | ||
| Collimator | ||||
| Type | : | Manual, with LED light field indicator | ||
| Rotation | : | ±120° | ||
| Filters | : | Built in filter disk for manual filter selection | ||
| : | No filter; 1 mm Al + 0.1 mm Cu; 1 mm Al + 0.2 mm Cu; 2 mm Al | |||
| Light indication when filter is selected | : | Yes | ||
| Manual SID indication | : | Yes | ||
| The Dose Area Product meter measures the X-ray dose output at the collimator and reports the | ||||
| Dose Area Product meter | : | measured Dose Area Product (μGy*m2 ) to the DICOM header of the image. | ||
| Dimensions (l x w x h) | : | 170 mm x 170 mm x 18 mm (6.7″ x 6.7″ x 0.7″) | ||
| Active area | : | 147 mm x 147 mm (5.8″ x 5.8″) | ||
| 4 | ELEVA WORKSPOT | 1 | ||
| Keyless entry | : | Yes, using otorized pasword | ||
| Hard disk | : | 1 TB | ||
| RAM storage capacity | : | 16 GB | ||
| Image storage | : | up to 4,000 images | ||
| Monitor | : | 17″-LCD color touch-screen monitor 1280 x 1024 at 60 Hz | ||
| Generator control | : | Integrated into Eleva software More than 600 pre-programmable APRs | ||
| 5 | SKYPLATE LARGE WIRELESS DETECTOR | 1 | ||
| Type | : | Digital flat detector Cesium Iodide (CsI) | ||
| Housing | : | Carbon fiber | ||
| Detector sizes | : | 35 cm x 43 cm (14″ x 17″) | ||
| Active area | : | 34.48 cm x 42.12 cm (13.7″ x 16.7″) | ||
| Image matrix size | : | 2330 x 2846 pixels | ||
| Detector pixels | : | 6.6 Megapixel | ||
| Image resolution | : | up to 3.38 Lp/mm | ||
| Pixel size | : | 148 μm | ||
| Weight (incl. battery) | : | large 2.8 kg (6.2 lbs) | ||
| Maximum patient weight | : | 100 kg (220 lbs) on 4 cm disk for weight bearing examinations | ||
| 300 kg (662 lb) for distributed load, e. g. chest examinations in bed | ||||
| 6 | SKYPLATE SMALL WIRELESS DETECTOR | 1 | ||
| Type | : | Digital flat detector Cesium Iodide (CsI) | ||
| Housing | : | Carbon fiber | ||
| Detector sizes | : | 24 cm x 30 cm | ||
| Active area | : | 22.3 cm x 28.4 cm | ||
| Image matrix size | : | 1500 x 1920 pixels | ||
| Detector pixels | : | 2.9 Mega pixels | ||
| Image resolution | : | up to 3.38 Lp/mm | ||
| Pixel size | : | 148 μm | ||
| Weight (incl. battery) | : | 1.6 kg including battery | ||
| Maximum patient weight | : | 100 kg (220 lbs) on 4 cm disk for weight bearing examinations | ||
| 300 kg (662 lb) for distributed load, e. g. chest examinations in bed | ||||
| Philips SkyFlow Plus is an intelligent software that produces images with grid-like contrast by managing | ||||
| 7 | SKYFLOW PLUS | : the effect of scattered radiation for non-grid exams across all anatomies. | 1 | |
| Provides time savings, since there is no need to attach/detach a grid | : | Yes | ||
| Obtain high-quality images equivalent to images acquired with a grid | : | Yes | ||
| Saves X-ray dose due to no retakes caused by grid misalignment | : | Yes | ||
| Works fully automatic and patient adaptive without special attention | : | Yes | ||
| UNIQUE 2 is the intuitive user interface provides an excellent opportunity to tailor the image to suit | ||||
| 8 | UNIQUE 2 | : your needs. image parameters can be easily adapted for each anatomy type, depending on the | 1 | |
| preferred focus. For ease of use, all settings can be customized through well-structured sliders within | ||||
| the interface | ||||
| Increased image contrast | : | Yes | ||
| Enhanced details | : | Yes | ||
| Reduced noise and artifacts | : | Yes | ||
| Optimal contrast harmonization with enhanced details | : | Yes | ||
| Analyze operator-rejected images and reasons for rejection with this powerful image statistic tool. | ||||
| 9 | CLINICAL QC | : Clinical QC also serves to monitor and analyze general parameters. Clinical QC provides valuable | 1 | |
| support for your department’s quality standards and for teaching situations. | ||||
| Easy and convenient monitoring of departmental quality standards | : | Yes | ||
| Dose documentation per image and examination | : | Yes | ||
| Presets of reasons for image rejection | : | Yes | ||
| Define time period statistic | : | Yes | ||
| • Statistic presentation as bar or pie chart at Eleva Workspot | : | Yes | ||
| 10 | DICOM | 1 | ||
| Dose Reporting in DICOM Structured Report format | : | Yes | ||
| DICOM WLM (Work List Management) | : | Yes | ||
| DICOM MPPS (Modality Performed Procedure Step) | : | Yes | ||
| DICOM Print | : | Yes | ||
| DICOM Image Export incl. Storage Commit | : | Yes | ||
| DICOM media | : | Yes | ||
| ACESSORIES | ||||
| 11 Wireless detector mobile holder | : | Yes | 1 | |
| 12 Wireless Remote Control | : | Yes | 1 | |
| 13 Detector handle for Large detector | : | Yes | 1 | |
| 14 Dry Printer | : | Yes | 1 | |
| 15 Mobile Shielding | : | Yes | 1 | |
| 16 Apron | : | Yes | 1 | |
| 17 Warranty (year) | : | Yes | 2 | |