| 1 | Table | Description | 1 | |
| Dimensions | ||||
| Table height | : 62 cm – 142 cm (24.4″ – 59.9″), elevating, motorized | |||
| Elevation range | : 80 cm (31.5″) | |||
| Elevation speed | : | 2.5 cm/s (1″/s) | ||
| Movements | ||||
| Tilt angle | : | – 90° to + 90° | ||
| Tilt speed | : | 2 speeds, 4.5°/s and 6.5°/s | ||
| Automatic stop | : In horizontal position | |||
| 2 | Table top | 1 | ||
| Dimensions | ||||
| Total dimensions | : | ±73.8 cm x 235.6 cm (29″ x 92.8″) | ||
| Radiolucent area | : | ±221.4 cm x 55.4 cm (87.2″ x 21.8″) | ||
| Table top to detector distance | : | Min. 10.2 cm (4″) | ||
| Material | : | Micro sandwich of laminate, carbon and foam | ||
| X-ray attenuation | : | 0.6 mm (0.02″) Al (at 100 kVp, HVL = 2.7 mm (0.1″) Al) | ||
| Maximum patient weight without limitations in movements | : | ±284 kg (626 lbs) | ||
| Movements | ||||
| Lateral | : ± 32 cm (12.6″) = (± 16 cm (6.3″)) | |||
| Longitudinal | : Only detector movements to improve patient comfort | |||
| Movement speed | ||||
| Lateral | : ±5 cm/s (1.9″/s), soft start and stop Auto centering | |||
| Longitudinal | : | Detector movement 3 cm/s – 20 cm/s (1.2″/s – 7.9″/s) | ||
| Tube Column-detector Assembly | ||||
| Movements | ||||
| Range | : ±160 cm (63″) longitudinal, motorized | |||
| Speed | : Variable, slow for positioning, high for travel maximum speed from 3 cm – 20 cm/s (1.2″/s – 7.9″/s) |
|||
| Patient coverage | : ±203 cm (79.9″), without patient movement | |||
| Angulation | ||||
| Movements | : | Motorized | ||
| Range | : | ± 40° | ||
| Speed | : | ±11.2°/s | ||
| Supporting functions | : | Automatic centering of target organ during oblique projections in fluoroscopy | ||
| : Oblique exposures at both ends of tabletop possible | ||||
| Source image distance | ||||
| Range | : | ±113 cm – 183 cm (44″ – 72″) | ||
| Movements | : | Motorized | ||
| Speed | : | ±4.1 cm/s (1.6″/s) | ||
| Focal spot to floor distance (in 90° position) | : | ±51.5 cm – 211.5 cm (20.3″ – 83.3″) (without angulation) | ||
| Tube rotation | : | Manual | ||
| Range | : – 90°/ + 180° | |||
| Yes, this usefull for driving the x-ray beam with AEC for easy positioning patient and guaranty there have adequate x-ray |
||||
| 3 | 7 Amplimat chamber on the fluoroscopy table | : | 1 | |
| dose for better image quality | ||||
| 4 | Compressor | 1 | ||
| Movements | : | Motorized | ||
| Activation | : | Remote controlled | ||
| Compression force | : | Variable, 3 kg – 15 kg (6.6 lbs – 33 lbs) in 0.5 kg (1.1 lbs) steps | ||
| Compressor parking | : | Automatic, behind tube stand | ||
| Cone | : Removable | |||
| 5 | Grid | : | Stationary Grid | 2 |
| Material | : | Carbon fiber | ||
| Lines | : | 44 lines/cm | ||
| Focus (FO) | : | ±120 cm (47.2″) and 180 cm (70.9″) | ||
| Automatic selection of Grid according to SID and programmed examination plus motorized removal of grids. Fully |
||||
| Multi grid system | : | automatic, no grid handling necessary | ||
| Trolley for in-room control of imaging and geometry parameters, including touchscreen geometry control console, |
||||
| 6 | Nearby control trolley | : footswitch for fluoroscopy and exposure and sealed waterproof keyboard with touchpad. | 1 | |
| 7 | X-ray generation | 1 | ||
| Exposure output power | : | 65 kW | ||
| kV | : | 40 – 150 kV (Second beam table and wall Bucky) | ||
| kV | : | 40 – 125 kV (dynamic exposures) | ||
| mA range | : | 1 – 900 mA | ||
| ms range | : | 1 ms – 4 s with AEC (Automatic Exposure Control) | ||
| ms range | : 1 ms – 16 s without AEC | |||
| Manual mode | ||||
| Two-factor technique (kV – mAs) | : Yes | |||
| Three-factor technique (kV – mA – s) | : Yes | |||
| Automatic mode | ||||
| One-factor falling load (kV) | : Yes | |||
| Two-factor constant load (kV/mA) | : Yes | |||
| Automatic kV reduction | : Yes | |||
| IQX provides excellent, reliable and consistent image quality for digital exposures, both in static and dynamic fluoroscopy |
||||
| studies. IQX controls and adapts the exposure parameters within the X-ray pulse. The automatic and fast regulation of kV |
||||
| IQX Intelligent exposur | : | during each exposure leads to high image quality for all types of studies, for all patients. | ||
| 8 | Tubes on Fluoroscopy table | 1 | ||
| Dual Focal spot | : | Yes | ||
| Min. anode speed | : | up to 10K revolutions/minute | ||
| Anode angle | : | ≥12° | ||
| Maximum voltage | : | ±110 to 125 kV | ||
| Nominal anode input power | : | 44 kW / 64 kW | ||
| Double tube overload protection | : | Yes | ||
| Total filtration minimum | : | 2.5 mm AI (IEC 60522, 75 kV) | ||
| 9 | 2nd Tube | 1 | ||
| Dual Focal spot | : | Yes (0.6 / 1.2) | ||
| Min. anode speed | : | up to 10K revolutions/minute | ||
| Anode angle | : | 13° | ||
| Maximum voltage | : | 150 kV | ||
| Nominal anode input power | : | 33 kW / 100 kW | ||
| Double tube overload protection | : | Yes | ||
| 10 Eleva Tube Head | 1 | |||
| Data displayed at the Eleva Tube Head | ||||
| Preview images | : | Yes | ||
| Grid status | : | Yes | ||
| Collimation field size | : | Yes | ||
| Angulation | : | Yes | ||
| Generator settings (kV, mA, mAs, AEC) | : | Yes | ||
| 11 Detectors (Fix on Fluoroscopy table) | 1 | |||
| Electronic flat detector (CsI/Cesium Iodide) | : | Yes | |
| Maximum field of view | : | 43 cm x 43 cm (17″ x 17″) | |
| Image matrix | : | 2880 x 2881 pixels | |
| Pixel size | : | 148 μm | |
| Bit depth | : | 16 bits | |
| Exposure dose | : | 0.1 – 10 μGy | |
| Cooling | : | Automatic internal air cooling system | |
| 12 SkyPlate (Wireless detector) | 1 | ||
| Type | : | Wireless, Digital CsI (Cesium Iodide) flat detector | |
| Housing material and Sensor protection material | : | Carbon fiber | |
| Detector size | : | 35 cm x 43 cm (14″ x 17″) | |
| Active area | : | 34.48 cm x 42.12 cm (13.6″ x 16.6″) | |
| Image matrix size | : | 2,330 x 2,846 pixel | |
| Pixel size | : | 148 μm | |
| A/D Conversion (bits) | : | 16 bits | |
| Weight (incl. battery) | : | 2.8 kg (6.2 lb) including battery | |
| 100 kg (220 lb) on 4 cm disk for weight bearing examinations, 300 kg (662 lb) for distributed load, e. g. chest examinations |
|||
| Max. patient weight | : | ||
| in bed | |||
| Dust/water protection | : IP43 | ||
| 13 Stitching Function | : Yes | 1 | |
| Capability doing Automatic image stitching, on the fluoroscopy table with paralel stitching technique. As for standing |
|||
| : (table tilt to standing 90 ° or patient stitching in vertical posistion also laying down on the fluoroscopy table and the |
|||
| stitching in horizontal projection) | |||
| 14 Eleva workspot | 1 | ||
| Processor Computer | : | Based on 3.9 GHz, Intel Core I7 | |
| Memory | : | 16 GB RAM | |
| Local storage | : | 1 TB Solid State Disk (SSD, 800 GB free for clinical images) | |
| Dynamic unique for fluoroscopy and radiography | : | Yes | |
| Multi image review | : | Mosaic view of 4, 9 or 16 single images | |
| Zoom 1 : 1, 2 : 1, 3 :3 | : | Yes | |
| Monitors | |||
| Two high quality color LCD monitors, one with touchscreen | : Yes | ||
| Size | : 21.3 inch | ||
| 15 UNIQUE2 Dynamic UNIQUE | : Yes | 1 | |
| 2nd generation of post processing software that automatically adjust contras and enhance image detail, optiming image |
|||
| processing that reducing motion blur (leg effec) and optimaling image from the first frame so no initial dose needed. |
|||
| Adjust contras and minimizing noise. | |||
| 16 DICOM | 1 | ||
| DICOM interoperability | : | Yes | |
| DICOM Print for Rad images (SCU) | : | Yes | |
| DICOM store (SCU) DICOM storage commit | : | Yes | |
| DICOM worklist management (SCU) | : | Yes | |
| DICOM MPPS (SCU) | : | Yes | |
| DICOM structured dose report | : Yes | ||
| 17 Monitor to be placed in examination room | 1 | ||
| Type | : | LCD color monitor IPS | |
| Size | : | 21.3″ | |
| Matrix | : | 1600″ x 1200″ (2 Megapixel) | |
| Luminance ratio | : | > 800:1 | |
| Dimensions | : approx. 495 x 425 mm (19.5″ x 16.7″) | ||
| Acessories | ||||
| 18 | 2nd WS (i7, 16GB DDR-2, 2TB, 27″ QHD) or Similar | : | Yes | 1 |
| 19 | Handgrips | : | Yes | 1 |
| 20 | Ceiling Construction | : | Yes | 1 |
| 21 | Dry Printer | : | Yes | 1 |
| 22 | Panel | : | Yes | 1 |
| 23 | UPS 3kVA | : | Yes | 1 |
| 24 | Apron | : | Yes | 1 |
| 25 | Warranty (year) | : | Yes | 3 |