No | Description | Qty | ||
1 | DIGITAL DIAGNOST C90 HIGH PERFORMANCE The high performance room configuration with fully motorized ceiling suspension and vertical movable stand including a fixed 43 cm x 43 cm (17″ x 17″) detector, is especially designed for high patient throughput. You can easily position the detector and X-ray tube close to the patient to perform all requested projections without relocating the patient. The configuration with the table-based SkyPlate tray, together with the large SkyPlate detector, allows for an exceptional variety of applications. Power consumption (at 400 V / 50 Hz) Stand by :340 W Off :100 W Temperature Operation :+18 °C to +30 °C Rel. humidity Operation :30 % to 75 % Air pressure :700 hPa to 1.060 hPa |
1 | ||
2 | CEILING SUSPENSION | 1 | ||
Four-part aluminium telescopic column with spring counter balanced holder for X-ray tube assembly; adaptable to | ||||
Column Type | : | individual room heights | ||
Ceiling height at source image distance 110 cm (44″) | : ± 2.84 m to 3.21 m (9′ 3.8″ to 10′ 6.3″) | |||
Movements | ||||
Longitudinal travel with Comfort Position | : | ± 3.26 m (10′ 8.3″) | ||
Transverse travel – short | : | ± 1.49 m (4′ 11″) | ||
Transverse travel – long | : | ± 3.21 m (10′ 6.4″) | ||
Vertical travel | : ± 1.65 m (5′ 5.2″) | |||
X-ray tube assembly | ||||
Minimum ceiling source distance | : | ± 87.1 cm (34.3″) | ||
Possible room height adjustment | : | ± 37.5 cm (14.8″) | ||
Lowest tube position with vertical stand | : | ± 30 cm (11.8″) measured from center of beam to the floor | ||
Lowest tube position with vertical movabable stand | : | ± 35 cm (13.8″) measured from center of beam to the floor | ||
Tube assembly rotation around vertical axis | : | 360° (±180°) with lock position every 45° | ||
Tube assembly rotation around horizontal axis | : 230° (± 115°) Servo supported movement, stop-lock positions at 0° and ±90° | |||
Rail system | ||||
Hardware | : Ceiling rail system made of anodized aluminium for long service life | |||
Length of rails | : ± 4.30 m (14′ 1.3″) | |||
Collimator | ||||
Type | : | Motorized automatic collimation or Manual overrule possible, with LED light field indicator | ||
Angle of rotation | : | ±45° | ||
Timer switch | : | up to 30 s | ||
Inherent filter value | : | Min. 0.1 mm Al at 75 kV | ||
Added filters to choose from | : | no filter; 1 mm Al + 0.1 mm Cu or 1 mm Al + 0.2 mm Cu | ||
: 2 mm Al + 0.5 mm Cu integrated filter only for detector callibration | ||||
3 | ELEVA TUBE HEAD | 1 | ||
Data displayed at the Eleva Tube Head | ||||
Patient data | : | Yes | ||
Live camera image | : | Yes | ||
Preview images | : | Yes | ||
Grid status | : | Yes | ||
Collimation field size | : | Yes | ||
Angulation | : | Yes | ||
Generator settings | : Yes | |||
4 | LIVE CAMERA PACKAGE | 1 | ||
Bird’s eye view for better visibility of collimation line | : | Yes | ||
IR cut filter for natural colors | : | Yes | ||
Highly light sensitive for low light environments | : | Yes | ||
Low latency for precise positioning | : | Yes | ||
Can be switched on and off at any time | : Yes | |||
Display of live camera image and collimation light at the Eleva Tube Head and at the Eleva Workspot | ||||
: Yes | ||||
5 | X-RAY GENERATION | 1 | ||
Generator | : | 65 kW | ||
Mains voltage | : | 380 V/400 V; 50 Hz, 3-phase and 480V; 60 Hz, 3-phase | ||
Max. mains resistance | : | ≤0.2 Ω at 380 V and ≤0.3 Ω at 400/480 V | ||
Max. mains current at 400 V or 380 V | : | 134 A or 150 A | ||
Max. tube voltage | : | 150 kV | ||
Max. tube current (at 80 kV) | : | 800 mA | ||
mAs product without AEC | : | 0.4 to 850 mAs | ||
mAs product with AEC | : | 0.5 to 600 mAs | ||
Compatible with VarioFocus | : | Yes | ||
Safety, Tube overload protection | : Yes | |||
6 | X-RAY TUBE | 1 | ||
Type | : | High power X-ray tube (SRO 33100) | ||
Two focal spots | : | 0.6 and 1.2 | ||
Max. power with focal spot 0.6 or with focal spot 1.2 | : | 33 kW / 100 kW | ||
Anode angle | : | 13° | ||
Max. tube voltage | : | 150 kV | ||
Anode heat storage capacity | : | 220 kJ (300 kHU) | ||
Continuous Input Power for Assembly | : | 250 W up to 3 hours | ||
Min. anode speed | : | 8,000 to 10,000 revolutions/minute | ||
Built in filter | : | 2 mm Al (0.079″) | ||
Compatible with VarioFocus | : | Yes | ||
Double tube overload protection | : Yes | |||
7 | FLAT DETECTOR (Fixed detector) | 1 | ||
Type | : | Digital CsI (Cesium Iodide) flat detector | ||
Detector size | : | 43 cm x 43 cm (17″ x 17″) | ||
Active area | : | 42 cm x 42.5 cm (16.5″ x 16.7″) | ||
Image matrix size | : | 2,840 pixel x 2,874 pixel | ||
Detector pixels | : | 8.2 Megapixel | ||
Pixel size | : | 148 μm | ||
Image resolution | : up to 3.4 Lp/mm | |||
8 | SKYPLATE DETECTOR (Wireless detector) | 1 | ||
Type | : Wireless, Digital CsI (Cesium Iodide) flat detector | |||
Housing material and Sensor protection material | : Carbon fiber | |||
Detector size | : 35 cm x 43 cm (14″ x 17″) |
Active area | : | 34.48 cm x 42.12 cm (13.6″ x 16.6″) | ||
Image matrix size | : | 2,330 x 2,846 pixel | ||
Detector pixels | : | 6.6 Megapixel | ||
Pixel size | : | 148 μm | ||
Image resolution | : | up to 3.38 Lp/mm | ||
Energy range (kVp) | : | 40 – 150 | ||
A/D Conversion (bits) | : | 16 | ||
Weight (incl. battery) | : | 2.8 kg (6.2 lb) incl. lead backer | ||
Max. patient weight | : | 100 kg (220 lb) on 4 cm disk for weight bearing examinations | ||
: | 300 kg (662 lb) for distributed load, e. g. chest examinations in bed | |||
Dust/water protection | : | IP43 | ||
General environmental conditions | ||||
Temperature range | : 10°C to 35°C | |||
Relative humidity range | : 0% to 80% | |||
Ambient pressure range | : 700 hPA to 1060 hPA | |||
Battery | ||||
Technology | : Exchangeable lithium ion battery | |||
Battery charging time | : ±4 hours max. for 100% charge | |||
Bar charge status color indication per battery | : 0-25%; 25-50%; 50-75%, 75-100 | |||
Electronic physical protection system with motor shutdown for downward movement; upward movement is possible at | ||||
9 | EXAMINATION TABLE | 1 | ||
any time | ||||
Type | : Height adjustable table | |||
Table base | : ± 51.5 cm to 91.5 cm (20.3″ to 36″) above floor, motorized adjustment | |||
Max. patient weight | : 375 kg (820 lb) | |||
Wide table top | ||||
Dimensions (l x w) | : ± 240 cm x 85 cm (7′ 10.5″ x 33.5″) | |||
longitudinal travel | : ±60 cm (±23.6″) | |||
transverse travel | : ±18 cm (±7.1″) | |||
10 | VERTICAL STAND | 1 | ||
Type | : | (VM) Vertical movable stand | ||
Hardware | : | Counterbalanced rugged column for motorized vertical movement of the detector unit | ||
Vertical travel | : | 35 cm to 185 cm (13.8″ to 6′ 08″), measured at center of detector | ||
Horizontal travel (motorized) | : 3.475 m (11′ 4.8″) | |||
Multi-purpose arm | ||||
Swiveling range | : | 0° to 90° (right or left orientated execution) | ||
Lock-in positions | : | manual or every 15° | ||
Tilt angle, horizontal axis | : | –20° to +90°, motorized tilting | ||
Tilt angle, vertical axis | : | +45° to –23°, manual tilting | ||
Automatic exposure control (AEC) | : | 5 AEC measuring fields | ||
Operating | : | 2 user interfaces (left and right) and wireless remote control | ||
Grips | : | Patient grips arranged on the left and the right of the detector unit | ||
Brakes | : All movements are locked when system is switched of | |||
Philips SkyFlow Plus is an intelligent software that produces images with grid-like contrast by managing the effect of | ||||
11 | SKYFLOW PLUS | scattered radiation for non-grid exams across all anatomies. | 1 | |
Provides time savings, since there is no need to attach/detach a grid | : | Yes | ||
Obtain high-quality images equivalent to images acquired with a grid | : | Yes | ||
Saves X-ray dose due to no retakes caused by grid misalignment | : | Yes | ||
Works fully automatic and patient adaptive without special attention | : Yes | |||
The automatic image stitching software is a dedicated orthopedic feature to automatically acquire long-length and | ||||
12 | AUTOMATIC IMAGE STITCHING | 1 | ||
: long-spine images | ||||
Ability to do stitching procedures with vertical stand and also on the table | : Yes | |||
Easy for the technologist to use by simply defining the collimation on the patient | : Yes | |||
System automatically acquires the number of images needed based on the defined collimation | ||||
: | Yes | |||
Automatic X-ray tube and detector movements during acquisition | : | Yes | ||
Software automatically stitches acquired images into one composite image | : | Yes | ||
Patient stand for streamlined patient positioning | : | Yes | ||
Single-focus X-ray tube rotation to reduce image distortion | : | Yes | ||
Dedicated orthopedic measurements included | : | Yes | ||
13 | UNIQUE 2 | : UNIQUE 2 is the intuitive user interface provides an excellent opportunity to tailor the image to suit your needs. image | 1 | |
parameters can be easily adapted for each anatomy type, depending on the preferred focus. | ||||
Increased image contrast | : Yes | |||
Enhanced details | : Yes | |||
Reduced noise and artifacts | : Yes | |||
Optimal contrast harmonization with enhanced details, while remains natural | : Yes | |||
Philips VarioFocus is a unique generator technology that provides outstanding image resolution by simultaneously | ||||
14 | VARIOFOCUS | : | using both the small and large tube filaments. | 1 |
Outstanding image quality through mixed focus spots adapted to each examination | : | Yes | ||
Outstanding resolution at the power level required | : | Yes | ||
Reduced exposure time | : | Yes | ||
Reduced motion artifacts | : | Yes | ||
Reduced geometrical blur | : | Yes | ||
Fully automatic | : | Yes | ||
Analyze operator-rejected images and reasons for rejection with this powerful image statistic tool. Clinical QC also | ||||
15 CLINICAL QC | serves to monitor and analyze general parameters. Clinical QC provides valuable support for your department’s quality 1 | |||
: standards and for teaching situations. | ||||
Easy and convenient monitoring of departmental quality standards | : | Yes | ||
Dose documentation per image and examination | : | Yes | ||
Presets of reasons for image rejection | : | Yes | ||
Define time period statistic | : | Yes | ||
Statistic presentation as bar or pie chart at Eleva Workspot | : Yes | |||
16 | ELEVA WORKSPOT | 1 | ||
Eleva Workspot compute | : | Eleva Workspot easily with the modern 2 Megapixel touch screen monitor | ||
Operating system | : | Windows 10 | ||
Processor | : | Intel® CoreTM i5 Processor | ||
Hard disc | : | 240 GB SSD | ||
Image storage | : | Typically around 10,000 images | ||
RAM storage capacity | : | 16 GB | ||
CD/DVD drive | : | CD reader/writer | ||
Keyboard with mouse and function buttons | : | Yes | ||
Display | : 21.3″ high quality LCD color touch screen |
17 | DICOM | 1 | ||
Dose Reporting in DICOM Structured Report format | : | Yes | ||
DICOM WLM (Work List Management) and Classic RIS | : | Yes | ||
DICOM MPPS (Modality Performed Procedure Step) | : | Yes | ||
DICOM Print | : | Yes | ||
DICOM Image Export incl. Storage Commit | : | Yes | ||
DICOM media on CD-R and DVD-R | : Yes | |||
ACESSORIES | ||||
18 COMPR. BELT SINGLES. PULL | : | Yes | 1 | |
19 DETECTOR MOBILE HOLDER | : | Yes | 1 | |
20 PAIR OF HANDGRIPS | : | Yes | 1 | |
21 STRETCH GRIP FOR VS/VM STAND | : | Yes | 1 | |
22 LARGE SKYPLATE HANDLE | : | Yes | 1 | |
23 GRID 40/12/110 (Bucky table) | : | Yes | 1 | |
24 GRID 40/8/140 (Bucky stand) | : | Yes | 1 | |
25 GRID 40/8/180 (Bucky stand) | : | Yes | 1 | |
26 WARRANTY (year) | : Yes | 2 |