1 | Table | Description | 1 | |
Dimensions | ||||
Table height | : 62 cm – 142 cm (24.4″ – 59.9″), elevating, motorized | |||
Elevation range | : 80 cm (31.5″) | |||
Elevation speed | : | 2.5 cm/s (1″/s) | ||
Movements | ||||
Tilt angle | : | – 90° to + 90° | ||
Tilt speed | : | 2 speeds, 4.5°/s and 6.5°/s | ||
Automatic stop | : In horizontal position | |||
2 | Table top | 1 | ||
Dimensions | ||||
Total dimensions | : | ±73.8 cm x 235.6 cm (29″ x 92.8″) | ||
Radiolucent area | : | ±221.4 cm x 55.4 cm (87.2″ x 21.8″) | ||
Table top to detector distance | : | Min. 10.2 cm (4″) | ||
Material | : | Micro sandwich of laminate, carbon and foam | ||
X-ray attenuation | : | 0.6 mm (0.02″) Al (at 100 kVp, HVL = 2.7 mm (0.1″) Al) | ||
Maximum patient weight without limitations in movements | : | ±284 kg (626 lbs) | ||
Movements | ||||
Lateral | : ± 32 cm (12.6″) = (± 16 cm (6.3″)) | |||
Longitudinal | : Only detector movements to improve patient comfort | |||
Movement speed | ||||
Lateral | : ±5 cm/s (1.9″/s), soft start and stop Auto centering | |||
Longitudinal | : | Detector movement 3 cm/s – 20 cm/s (1.2″/s – 7.9″/s) | ||
Tube Column-detector Assembly | ||||
Movements | ||||
Range | : ±160 cm (63″) longitudinal, motorized | |||
Speed | : Variable, slow for positioning, high for travel maximum speed from 3 cm – 20 cm/s (1.2″/s – 7.9″/s) |
|||
Patient coverage | : ±203 cm (79.9″), without patient movement | |||
Angulation | ||||
Movements | : | Motorized | ||
Range | : | ± 40° | ||
Speed | : | ±11.2°/s | ||
Supporting functions | : | Automatic centering of target organ during oblique projections in fluoroscopy | ||
: Oblique exposures at both ends of tabletop possible | ||||
Source image distance | ||||
Range | : | ±113 cm – 183 cm (44″ – 72″) | ||
Movements | : | Motorized | ||
Speed | : | ±4.1 cm/s (1.6″/s) | ||
Focal spot to floor distance (in 90° position) | : | ±51.5 cm – 211.5 cm (20.3″ – 83.3″) (without angulation) | ||
Tube rotation | : | Manual | ||
Range | : – 90°/ + 180° | |||
Yes, this usefull for driving the x-ray beam with AEC for easy positioning patient and guaranty there have adequate x-ray |
||||
3 | 7 Amplimat chamber on the fluoroscopy table | : | 1 | |
dose for better image quality | ||||
4 | Compressor | 1 | ||
Movements | : | Motorized | ||
Activation | : | Remote controlled | ||
Compression force | : | Variable, 3 kg – 15 kg (6.6 lbs – 33 lbs) in 0.5 kg (1.1 lbs) steps | ||
Compressor parking | : | Automatic, behind tube stand | ||
Cone | : Removable |
5 | Grid | : | Stationary Grid | 2 |
Material | : | Carbon fiber | ||
Lines | : | 44 lines/cm | ||
Focus (FO) | : | ±120 cm (47.2″) and 180 cm (70.9″) | ||
Automatic selection of Grid according to SID and programmed examination plus motorized removal of grids. Fully |
||||
Multi grid system | : | automatic, no grid handling necessary | ||
Trolley for in-room control of imaging and geometry parameters, including touchscreen geometry control console, |
||||
6 | Nearby control trolley | : footswitch for fluoroscopy and exposure and sealed waterproof keyboard with touchpad. | 1 | |
7 | X-ray generation | 1 | ||
Exposure output power | : | 65 kW | ||
kV | : | 40 – 150 kV (Second beam table and wall Bucky) | ||
kV | : | 40 – 125 kV (dynamic exposures) | ||
mA range | : | 1 – 900 mA | ||
ms range | : | 1 ms – 4 s with AEC (Automatic Exposure Control) | ||
ms range | : 1 ms – 16 s without AEC | |||
Manual mode | ||||
Two-factor technique (kV – mAs) | : Yes | |||
Three-factor technique (kV – mA – s) | : Yes | |||
Automatic mode | ||||
One-factor falling load (kV) | : Yes | |||
Two-factor constant load (kV/mA) | : Yes | |||
Automatic kV reduction | : Yes | |||
IQX provides excellent, reliable and consistent image quality for digital exposures, both in static and dynamic fluoroscopy |
||||
studies. IQX controls and adapts the exposure parameters within the X-ray pulse. The automatic and fast regulation of kV |
||||
IQX Intelligent exposur | : | during each exposure leads to high image quality for all types of studies, for all patients. | ||
8 | Tubes on Fluoroscopy table | 1 | ||
Dual Focal spot | : | Yes | ||
Min. anode speed | : | up to 10K revolutions/minute | ||
Anode angle | : | ≥12° | ||
Maximum voltage | : | ±110 to 125 kV | ||
Nominal anode input power | : | 44 kW / 64 kW | ||
Double tube overload protection | : | Yes | ||
Total filtration minimum | : | 2.5 mm AI (IEC 60522, 75 kV) | ||
9 | 2nd Tube | 1 | ||
Dual Focal spot | : | Yes (0.6 / 1.2) | ||
Min. anode speed | : | up to 10K revolutions/minute | ||
Anode angle | : | 13° | ||
Maximum voltage | : | 150 kV | ||
Nominal anode input power | : | 33 kW / 100 kW | ||
Double tube overload protection | : | Yes | ||
10 Eleva Tube Head | 1 | |||
Data displayed at the Eleva Tube Head | ||||
Preview images | : | Yes | ||
Grid status | : | Yes | ||
Collimation field size | : | Yes | ||
Angulation | : | Yes | ||
Generator settings (kV, mA, mAs, AEC) | : | Yes | ||
11 Detectors (Fix on Fluoroscopy table) | 1 |
Electronic flat detector (CsI/Cesium Iodide) | : | Yes | |
Maximum field of view | : | 43 cm x 43 cm (17″ x 17″) | |
Image matrix | : | 2880 x 2881 pixels | |
Pixel size | : | 148 μm | |
Bit depth | : | 16 bits | |
Exposure dose | : | 0.1 – 10 μGy | |
Cooling | : | Automatic internal air cooling system | |
12 SkyPlate (Wireless detector) | 1 | ||
Type | : | Wireless, Digital CsI (Cesium Iodide) flat detector | |
Housing material and Sensor protection material | : | Carbon fiber | |
Detector size | : | 35 cm x 43 cm (14″ x 17″) | |
Active area | : | 34.48 cm x 42.12 cm (13.6″ x 16.6″) | |
Image matrix size | : | 2,330 x 2,846 pixel | |
Pixel size | : | 148 μm | |
A/D Conversion (bits) | : | 16 bits | |
Weight (incl. battery) | : | 2.8 kg (6.2 lb) including battery | |
100 kg (220 lb) on 4 cm disk for weight bearing examinations, 300 kg (662 lb) for distributed load, e. g. chest examinations |
|||
Max. patient weight | : | ||
in bed | |||
Dust/water protection | : IP43 | ||
13 Stitching Function | : Yes | 1 | |
Capability doing Automatic image stitching, on the fluoroscopy table with paralel stitching technique. As for standing |
|||
: (table tilt to standing 90 ° or patient stitching in vertical posistion also laying down on the fluoroscopy table and the |
|||
stitching in horizontal projection) | |||
14 Eleva workspot | 1 | ||
Processor Computer | : | Based on 3.9 GHz, Intel Core I7 | |
Memory | : | 16 GB RAM | |
Local storage | : | 1 TB Solid State Disk (SSD, 800 GB free for clinical images) | |
Dynamic unique for fluoroscopy and radiography | : | Yes | |
Multi image review | : | Mosaic view of 4, 9 or 16 single images | |
Zoom 1 : 1, 2 : 1, 3 :3 | : | Yes | |
Monitors | |||
Two high quality color LCD monitors, one with touchscreen | : Yes | ||
Size | : 21.3 inch | ||
15 UNIQUE2 Dynamic UNIQUE | : Yes | 1 | |
2nd generation of post processing software that automatically adjust contras and enhance image detail, optiming image |
|||
processing that reducing motion blur (leg effec) and optimaling image from the first frame so no initial dose needed. |
|||
Adjust contras and minimizing noise. | |||
16 DICOM | 1 | ||
DICOM interoperability | : | Yes | |
DICOM Print for Rad images (SCU) | : | Yes | |
DICOM store (SCU) DICOM storage commit | : | Yes | |
DICOM worklist management (SCU) | : | Yes | |
DICOM MPPS (SCU) | : | Yes | |
DICOM structured dose report | : Yes | ||
17 Monitor to be placed in examination room | 1 | ||
Type | : | LCD color monitor IPS | |
Size | : | 21.3″ | |
Matrix | : | 1600″ x 1200″ (2 Megapixel) | |
Luminance ratio | : | > 800:1 | |
Dimensions | : approx. 495 x 425 mm (19.5″ x 16.7″) |
Acessories | ||||
18 | 2nd WS (i7, 16GB DDR-2, 2TB, 27″ QHD) or Similar | : | Yes | 1 |
19 | Handgrips | : | Yes | 1 |
20 | Ceiling Construction | : | Yes | 1 |
21 | Dry Printer | : | Yes | 1 |
22 | Panel | : | Yes | 1 |
23 | UPS 3kVA | : | Yes | 1 |
24 | Apron | : | Yes | 1 |
25 | Warranty (year) | : | Yes | 3 |