No | Philips CombiDiagnost R90 HP with Dual Tube | Qty | |||
1 | Table | Description | 1 | ||
Dimensions | |||||
Table height | : 62 cm – 142 cm (24.4″ – 59.9″), elevating, motorized | ||||
Elevation range | : 80 cm (31.5″) | ||||
Elevation speed | : | 2.5 cm/s (1″/s) | |||
Movements | |||||
Tilt angle | : | – 90° to + 90° | |||
Tilt speed | : | 2 speeds, 4.5°/s and 6.5°/s | |||
Automatic stop | : In horizontal position | ||||
2 | Table top | 1 | |||
Dimensions | |||||
Total dimensions | : | ±73.8 cm x 235.6 cm (29″ x 92.8″) | |||
Radiolucent area | : | ±221.4 cm x 55.4 cm (87.2″ x 21.8″) | |||
Table top to detector distance | : | Min. 10.2 cm (4″) | |||
Material | : | Micro sandwich of laminate, carbon and foam | |||
X-ray attenuation | : | 0.6 mm (0.02″) Al (at 100 kVp, HVL = 2.7 mm (0.1″) Al) | |||
Maximum patient weight without limitations in movements | : | ±284 kg (626 lbs) | |||
Movements | |||||
Lateral | : ± 32 cm (12.6″) = (± 16 cm (6.3″)) | ||||
Longitudinal | : Only detector movements to improve patient comfort | ||||
Movement speed | |||||
Lateral | : ±5 cm/s (1.9″/s), soft start and stop Auto centering | ||||
Longitudinal | : | Detector movement 3 cm/s – 20 cm/s (1.2″/s – 7.9″/s) | |||
Tube Column-detector Assembly | |||||
Movements | |||||
Range | : ±160 cm (63″) longitudinal, motorized | ||||
Speed | : Variable, slow for positioning, high for travel maximum speed from 3 cm – 20 cm/s (1.2″/s – 7.9″/s) | ||||
Patient coverage | : ±203 cm (79.9″), without patient movement | ||||
Angulation | |||||
Movements | : | Motorized | |||
Range | : | ± 40° | |||
Speed | : | ±11.2°/s | |||
Supporting functions | : | Automatic centering of target organ during oblique projections in fluoroscopy | |||
: Oblique exposures at both ends of tabletop possible | |||||
Source image distance | |||||
Range | : | ±113 cm – 183 cm (44″ – 72″) | |||
Movements | : | Motorized | |||
Speed | : | ±4.1 cm/s (1.6″/s) | |||
Focal spot to floor distance (in 90° position) | : | ±51.5 cm – 211.5 cm (20.3″ – 83.3″) (without angulation) | |||
Tube rotation | : | Manual | |||
Range | : | – 90°/ + 180° | |||
Stop position | : – 90°/ – 50°/ – 40°/ 0°/ + 40°/ + 50°/ + 90°/ + 180° | ||||
Yes, this usefull for driving the x-ray beam with AEC for easy positioning patient and guaranty there have adequate x-ray | |||||
3 | 7 Amplimat chamber on the fluoroscopy table | : | 1 | ||
dose for better image quality | |||||
4 | Compressor | 1 | |||
Movements | : | Motorized | |||
Activation | : | Remote controlled | |||
Compression force | : | Variable, 3 kg – 15 kg (6.6 lbs – 33 lbs) in 0.5 kg (1.1 lbs) steps | |||
Compressor parking | : Automatic, behind tube stand | ||||
Cone | : Removable | ||||
5 | Grid | : | Stationary Grid | 2 | |
Material | : | Carbon fiber | |||
Lines | : | 44 lines/cm | |||
Focus (FO) | : | ±120 cm (47.2″) and 180 cm (70.9″) | |||
Automatic selection of Grid according to SID and programmed examination plus motorized removal of grids. Fully | |||||
Multi grid system | : | automatic, no grid handling necessary | |||
Trolley for in-room control of imaging and geometry parameters, including touchscreen geometry control console, | |||||
6 | Nearby control trolley | : footswitch for fluoroscopy and exposure and sealed waterproof keyboard with touchpad. | 1 | ||
7 | X-ray generation | 1 | |||
Exposure output power | : | 80 kW | |||
kV | : | 40 – 150 kV (Second beam table and wall Bucky) | |||
kV | : | 40 – 125 kV (dynamic exposures) | |||
mA range | : | 1 – 1100 mA | |||
ms range | : | 1 ms – 4 s with AEC (Automatic Exposure Control) | |||
ms range | : 1 ms – 16 s without AEC | ||||
Manual mode | |||||
Two-factor technique (kV – mAs) | : Yes | ||||
Three-factor technique (kV – mA – s) | : Yes | ||||
Automatic mode | |||||
One-factor falling load (kV) | : Yes | ||||
Two-factor constant load (kV/mA) | : Yes | ||||
Automatic kV reduction | : Yes | ||||
IQX provides excellent, reliable and consistent image quality for digital exposures, both in static and dynamic fluoroscopy | |||||
studies. IQX controls and adapts the exposure parameters within the X-ray pulse. The automatic and fast regulation of kV | |||||
IQX Intelligent exposur | : | during each exposure leads to high image quality for all types of studies, for all patients. | |||
8 | Tubes on Fluoroscopy table | 1 | |||
Dual Focal spot | : | Yes | |||
Min. anode speed | : | up to 10K revolutions/minute | |||
Anode angle | : | ≥12° | |||
Maximum voltage | : | ±110 to 125 kV | |||
Nominal anode input power | : | 44 kW / 64 kW | |||
Double tube overload protection | : | Yes | |||
Total filtration minimum | : | 2.5 mm AI (IEC 60522, 75 kV) | |||
9 | 2nd Tube | 1 | |||
Dual Focal spot | : | Yes (0.6 / 1.2) | |||
Min. anode speed | : | up to 10K revolutions/minute | |||
Anode angle | : | ≥12° | |||
Maximum voltage | : | 150 kV | |||
Nominal anode input power | : | 44 kW / 64 kW | |||
Double tube overload protection | : | Yes | |||
10 Eleva Tube Head | 1 | ||||
Data displayed at the Eleva Tube Head | |||||
Preview images | : | Yes | |||
Grid status | : | Yes | |||
Collimation field size | : | Yes | |||
Angulation | : | Yes | |||
Generator settings (kV, mA, mAs, AEC) | : | Yes | |||
11 Fix Vertical Stand | 1 | ||||
Type | : | (VS) Fix Vertical Stand | |||
Vertical travel | : | 35 cm to 185 cm, measured at center of detector | |||
Automatic exposure control (AEC) | : | 5 AEC measuring fields | |||
Operating | : | 2 user interfaces (left and right) and wireless remote control | |||
Grips | : | Patient grips arranged on the left and the right of the detector unit | |||
Brakes | : All movements are locked when system is switched off | ||||
12 Detectors (Fix on Fluoroscopy table) | 1 | ||||
Electronic flat detector (CsI/Cesium Iodide) | : | Yes | |||
Maximum field of view | : | 43 cm x 43 cm (17″ x 17″) | |||
Image matrix | : | 2880 x 2881 pixels | |||
Pixel size | : | 148 μm | |||
Bit depth | : | 16 bits | |||
Exposure dose | : | 0.1 – 10 μGy | |||
Cooling | : | Automatic internal air cooling system | |||
13 SkyPlate (Wireless detector) | 1 | ||||
Type | : | Wireless, Digital CsI (Cesium Iodide) flat detector | |||
Housing material and Sensor protection material | : | Carbon fiber | |||
Detector size | : | 35 cm x 43 cm (14″ x 17″) | |||
Active area | : | 34.48 cm x 42.12 cm (13.6″ x 16.6″) | |||
Image matrix size | : | 2,330 x 2,846 pixel | |||
Pixel size | : | 148 μm | |||
A/D Conversion (bits) | : | 16 bits | |||
Weight (incl. battery) | : | 2.8 kg (6.2 lb) including battery | |||
100 kg (220 lb) on 4 cm disk for weight bearing examinations, 300 kg (662 lb) for distributed load, e. g. chest examinations | |||||
Max. patient weight | : | ||||
in bed | |||||
Dust/water protection | : Yes | ||||
14 Stitching Function | : Yes | 1 | |||
Capability doing Automatic image stitching, on the fluoroscopy table with paralel stitching technique. As for standing | |||||
: (table tilt to standing 90 ° or patient stitching in vertical posistion also laying down on the fluoroscopy table and the | |||||
stitching in horizontal projection) | |||||
15 Eleva workspot | 1 | ||||
Processor Computer | : | Based on 3.9 GHz, Intel Core I7 | |||
Memory | : | 16 GB RAM | |||
Local storage | : | 1 TB Solid State Disk (SSD, 800 GB free for clinical images) | |||
Dynamic unique for fluoroscopy and radiography | : | Yes | |||
Multi image review | : | Mosaic view of 4, 9 or 16 single images | |||
Zoom 1 : 1, 2 : 1, 3 :3 | : | Yes | |||
Monitors | |||||
Two high quality color LCD monitors, one with touchscreen | : Yes | ||||
Size | : 21.3 inch | ||||
16 UNIQUE2 Dynamic UNIQUE | : Yes | 1 | |||
2nd generation of post processing software that automatically adjust contras and enhance image detail, optiming image | |||||
processing that reducing motion blur (leg effec) and optimaling image from the first frame so no initial dose needed. | |||||
Adjust contras and minimizing noise. | |||||
17 DICOM | 1 | ||||
DICOM interoperability | : | Yes | |||
DICOM Print for Rad images (SCU) | : | Yes | |||
DICOM store (SCU) DICOM storage commit | : | Yes | |||
DICOM worklist management (SCU) | : | Yes | |||
DICOM MPPS (SCU) | : Yes | ||||
DICOM structured dose report | : Yes | ||||
18 Monitor to be placed in examination room | 1 | ||||
Type | : | LCD color monitor IPS | |||
Size | : | 21.3″ | |||
Matrix | : | 1600″ x 1200″ (2 Megapixel) | |||
Luminance ratio | : | > 800:1 | |||
Dimensions | : | approx. 495 x 425 mm (19.5″ x 16.7″) | |||
Weight | : approx. 7.1 kg (15.6 lbs) | ||||
Acessories | |||||
19 | 2nd WS (i7, 16GB DDR-2, 2TB, 27″ QHD) or Similar | : | Yes | 1 | |
20 | Handgrips | : | Yes | 1 | |
21 | Shoulder support | : | Yes | 1 | |
22 | Pair of leg supports | : | Yes | 1 | |
23 | Ceiling Construction | : | Yes | 1 | |
24 | Dry Printer | : | Yes | 1 | |
25 | Panel | : | Yes | 1 | |
26 | UPS 3kVA | : | Yes | 1 | |
27 | Apron | : | Yes | 1 | |
28 | Warranty (year) | : Yes | 3 |
- Beranda
- E-Katalog
- Imaging Systems
- Diagnost Radiography
- Philips CombiDiagnost R90 HP with Dual Tube
Philips CombiDiagnost R90 HP with Dual Tube
01 Overview
CombiDiagnost R90 is a remote controlled fluoroscopy system in combination with high-end digital radiography, designed to improve room utilization in a cost effective manner. A fully digital workflow, UNIQUE image quality and excellent dose management make the versatile system suitable for a wide range of examinations from pediatric to bariatric imaging.
Name: CombiDiagnost R90.
Brand: Philips
Certificate of Origin: Germany / Netherland
No. AKL: KEMENKES RI AKL 21501120698
- Price:
Rp.9.900.000.000,00
- E-Katalog LKPP Link: https://bit.ly/3VBJ3NQ
- Categories: Imaging Systems, Diagnost Radiography
- Share: