SPECIFICATION PHILIPS Azurion 7 M12/M20 and IntraSight IVUS with iFR-FFR plus SyncVision | ||
REFERENCE OF TECHNICAL | ||
Azurion 7 M12/M20 with Clarity IQ | ||
Gantry | ||
Gantry Mounting system | Ceiling Mounted | |
Longitudinal movement | motorized and manual range of 260 cm | |
The gantry positions can be recalled based on the stored image information |
Yes | |
The sequence of projections must be physician dependent |
Yes | |
Collision protection | Yes, With Body Guard | |
Patient support | ||
Floor-mounted radio translucent floating top. It shall be able to cover full head to toe examination without repositioning the patient |
Yes | |
The patient can have CPR at any position of the tabletop | Yes | |
Table Length | 319 cm | |
Table width | 50 cm | |
Pivot Table | -90°/+180° or -180°/90° | |
Generator | ||
Voltage range | 40-125 kV | |
Maximum current | 1000 mA | |
Different exposure protocols can be customized | Yes | |
Digital imaging and postprocessing | ||
DSA capability | Yes | |
ECG waveform can be recorded, stored, and displayed on the monitor with the image information? |
Yes | |
Detector | ||
Detector size | Available in two options, diagonal size 20 inch (M20) for large detector or diagonal size 12 inch (M12) for small detector. |
|
*Pembeli wajib menyebutkan ukuran detektor yang ingin dibeli pada saat proses pembelian melalui situs e-catalogue | ||
Monitors | ||
Monitor Type | Medical Grade | |
Monitor in Examination Room | – | 1 x 58-inch Medical Grade Monitor |
– | Equipped with PRO Features | |
Equipped with patient tableside touch screen module | Yes | |
Touch screen in patient tableside capability | – enhance image navigation on the TSM | |
– intuitive control of shutters and wedges by simply dragging | ||
the lines shown on top of the image |
– able to view images in the touch screen | ||
– provides intuitive zooming and panning functionality (also during fluoroscopy) |
||
– able to act as additional keyboard and mouse inside the control room |
||
Monitor in Control Room | – | Minimum 2 x 27-inch monitor |
– | Equipped with FlexSpot Features | |
Automatic positioning stand program | Yes | |
Automatic positioning stand program able to perform sequences of saved geometry position with a push of a button |
Yes | |
System functionality | ||
QA packages for Coronary analysis. | Yes | |
QA packages for Vascular analysis. | Yes | |
QA packages for Left Ventricular analysis. | Yes | |
Rotational Angiography (single Axis Rotation Scan) | Acquisition mode: Shall be possible for any body part to provide 3D impression of the complex vasculature and coronary arteries including lower limb. |
|
Rotation Swing / Simultaneous Dual Axis Rotation Scan: | Yes, With CardiacSwing | |
Yes, Interventional workstation for reconstructing images | ||
from Rotational Angiography to provide 3D images of brain & | ||
3D Rotational Angiography: | peripheral vessels to facilitate endovascular interventions in | |
neuro & vascular radiology. | ||
Pre- and post Stent Deployment Tool: | Yes, With StentBoost | |
Digital Subtracted Angiography | Yes | |
Subtracted Bolus Chase and Bolus Chase Reconstruction | Yes | |
Automatic creation and storage of a dynamic roadmap from each acquired coronary angiogram (Roadmap on coronary angiogram) |
Yes, Dynamic Coronary Roadmap | |
Capability to visualize of balloons and stents in coronary arteries on-the-fly to clarify the situation at any moment during an intervention (Live images) |
Yes, with StentBoost Live | |
Equipped with EP Navigator | Yes | |
Dose-saving measures | ||
Zero Dose Positioning | Yes | |
The anti-scatter grid shall be removable, e.g., for pediatric acquisitions |
Yes | |
Accumulated AK indication for different patient body zones, with warning in case 2Gy has been reached for every individual zone |
Yes | |
On screen display for stand position, tube load status, fluoro mode detector field size, copper filtration. Both rate and accumulated dose shall be available in the exam and control room. |
Yes |
Dose Reduction Technology | |
Dose reduction features | Yes, included |
Image harmonization | Image harmonization technology to avoid over radiated lung fields and dark spine areas |
Pre-filtration for different Fluoroscopy mode | Fluoroscopy mode of at least 3 selections at the table side of different pre-programmed dose rate and copper pre- filtration. |
Hemodynamic System | |
Flexibility mounting at the table side to improve access to the patient |
Yes |
Minimum monitoring capabilities: | |
• NIBP | Yes |
• SpO2 | Yes |
• Invasive Pressures | Yes |
• ECG | Yes |
Additional Workstation | |
Cardiac Image Review Workstation with minimum 1 concurrent user |
Yes |
Accessories | |
Intercom | Yes |
Peripheral X-ray Filter | Yes |
Arm Support | Yes |
Head Support | Yes |
Table mounted radiation shield | Yes |
Ceiling mounted radiation shield | Yes |
Exam Light | Yes |
Pan Handle | Yes |
Radiation protection sets | |
Lead Glass | Yes, 1 unit |
Lead Apron | Yes, 5 units |
Lead Goggle | Yes, 5 units |
Lead Thyroid Collar | Yes, 5 units |
IntraSight IVUS with iFR-FFR plus SyncVision | |
An Integrated IVUS imaging, physiology and co-registration Interventional applications platform that provides | |
enhancement and quantification of angiographic images during PCI. A real-time image processing system used in | |
diagnostic cardiac catheterization procedures and PCI. It combines iFR and IVUS data with the angiogram using iFR Co- | |
registration, IVUS Co-registration and Tri-registration features when used with Intrasight. |
Processing and data storage | |
Memory | Minimum 32GB SD RAM |
Hard drive capacity | Minimum 1TB SSD SATA |
Digital archiving capacity | Minimum Local, DVD, DICOM Network (includes Worklist management, DICOM Store) |
IntraSight interventional applications platform | |
• Intrasight CPU | Yes, 1 unit |
• Minimum 15-19” monitor kit | Yes, 1 unit |
• Mouse | Yes, 1 unit |
• Keyboard | Yes, 1 unit |
• Cabling kit | Yes, 1 unit |
• Touch screen module (TSM) | Yes, 1 unit |
Imaging (IVUS) license | |
Intravascular Ultrasound (IVUS) therapy using digital or rotational catheters |
Yes |
Offers the plug-and-play simplicity of digital IVUS | Yes |
Compatible with digital catheters that enable maximum penetration diameter of 60mm |
Yes |
ChromaFlo highlights blood flow red for easy assessment of stent apposition |
Yes |
Physiology license | |
Includes instantaneous Wave-free ratio (iFR) with hyperemia independent ability to assess serial lesions |
Yes |
Fractional Flow Reserve measurement measure ratio represents the potential decrease in coronary flow distal to the coronary stenosis |
Yes |
iFR Scout pullback shows most significant gradient in the mid-vessel lesion with diffuse proximal disease |
Yes |
SyncVision | |
SyncVision parts and accessories | |
•Syncvision Software and CPU | Yes, 1 unit |
•Monitor | Yes, 1 unit |
•Joystick | Yes, 1 unit |
•Mouse | Yes, 1 unit |
•Keyboard | Yes, 1 unit |
•Cabling kit | Yes, 1 unit |
iFR Co-registration | |
Provides localization of physiology measurements on the | Yes |
angiogram during the decide and confirm stages of a PCI | |
procedure |
IVUS Co-registration | |
Provides localization of the IVUS image on the angiogram during the decide, guide, and confirm stages of a PCI procedure |
Yes |
Physiology-IVUS Tri-registration | |
Obtain easy length measurements that combine IVUS and iFR information with the angiogram to help determine if a stent will meet the procedural objectives. |
Yes |
Device Detection | |
During live X-ray, in both fluoro and cine modes, this | Yes |
feature automatically and continuously generates a live, | |
stabilized, and enhanced image of stents and | |
bioresorbable vascular scaffolds, during device | |
positioning, balloon inflation and deployment of the | |
device. | |
Quantitative Coronary Analysis | |
Enables the calculation of an advanced QCA and vessel enhancement, just with one click on the vessel |
Yes |
Vessel Enhancement | |
Provide more clarity and enhanced image when visualizing the vessel. |
Yes |
Warranty for Philips main unit | 2 Years |